400-050-3910

The SINZEN Minfrared gas sensor is mainly composed of infrared correlation filter technology (GFC) and self-designed long optical path gas absorption cell (L-Cell), wavelength filter, infrared detector and high-precision signal processing circuit to complete the gas quantitative analysis in the infrared band. The sensor has the characteristics of high precision, good stability and fast response time. The product is mainly used in the following applications: Air quality testing; Industrial process gas analysis, etc.
Model
XZ-WM-003
When the infrared light passes through the measured gas cell, these gas molecules absorb the infrared light of a specific wavelength, and the absorption relationship obeys the Lambert-Beer absorption law. By measuring the light intensity of the absorption window and the non-absorption window, it is able to calculate the concentration of the measured gas. Imported infrared light source provides highly stable infrared light, which can meet the gas measurement of infrared gas absorption cell; highly sensitive detector cooperates with highly customized two filters with strong selectivity, select specific wavelength light to irradiate to the detector, so as to output voltage waveform with concentration signal, provide microprocessor for collection, calculation and calibration, and obtain concentration signal.
| Measuring principle | Infrared absorption(NDIR) |
| measuring gas | CO,CO2,CH4 |
| Measurement range | 0~10/50(500/2000)ppm,other ranges can be customized. |
| Sample gas flow | 1 L/min±10% |
| Response time | ≤64s |
| Indication error | ≤2%FS |
| Drift | ≤±1%FS/24h |
| Warm-up time | ≤60min |
| Output Interface | RS-232/RS-485/4-20mA |